Twiddler Syndrome without Lead Dislodgment Discovered by Remote Monitoring

Champagne, Catherine and Dognin, Nicolas and Molin, Franck and Steinberg, Christian and Philippon, François and Sarrazin, Jean-François and O'hara, Gilles and Nault, Isabelle and Roy, Karine and Plourde, Benoit and Blier, Louis and Champagne, Jean and Snyder, Christopher S. (2021) Twiddler Syndrome without Lead Dislodgment Discovered by Remote Monitoring. Case Reports in Cardiology, 2021. pp. 1-4. ISSN 2090-6404

[thumbnail of 8816524.pdf] Text
8816524.pdf - Published Version

Download (1MB)

Abstract

Twiddler syndrome is an uncommon yet dangerous phenomenon usually resulting in lead displacement with loss of capture of
cardiac implantable electronic devices. In this case report, we present an interesting case of Twiddler syndrome without lead
dislodgment which was detected by an alert triggered by an increase in impedance on remote monitoring